Table 11-33 DIN 11851 [bar]
PN (bar) / DN
Temperature TS (°C)
-10
0
50
100
140
25 / 50-100
25
25
25
25
25
40 / 10-40
40
40
40
40
40
Table 11-34 DIN 32676 & ISO 2852 [bar]
PN (bar) / DN
Temperature TS (°C)
-10
0
50
100
140
10 / 85-219.1
10
10
10
10
10
16 / 48.3-76.2
16
16
16
16
16
25 / 6.35-42.4
25
25
25
25
25
Note
Test pressure
Maximum allowable test pressure (MATP) for the flowmeter and process connection is 1.5
times the nominal pressure up to 150 bar (2176 psi).
11.13.2
Pressure - temperature ratings (nickel alloy sensors)
Figure 11-7
Nickel alloy: Metric flange ratings, EN 1092-1 (P: Process pressure; T: Process
temperature)
Technical data
11.13 Pressure - temperature ratings
SITRANS FC330
214
Operating Instructions, 01/2019, A5E44030648-AB
Содержание SITRANS FC330
Страница 14: ...Introduction 1 8 Notes on warranty SITRANS FC330 14 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 56: ...Installing mounting 4 3 Disassembly SITRANS FC330 56 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 126: ...Operating 7 2 Remote operation SITRANS FC330 126 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 158: ...Parameter assignment 8 2 Functions SITRANS FC330 158 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 216: ...Technical data 11 13 Pressure temperature ratings SITRANS FC330 216 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 224: ...Dimension drawings 12 7 Mounting bracket dimensions SITRANS FC330 224 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 236: ...HART communication B 1 Mode of operation HART function SITRANS FC330 236 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 274: ...Zero point adjustment SITRANS FC330 274 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 280: ...Index SITRANS FC330 280 Operating Instructions 01 2019 A5E44030648 AB ...