FSO (sensors
max. flow rate)
Error in % of actual mass flow rate with 95% confidence
(probability)
Actual mass
flow rate
-1.0
-0.5
1.0
0.5
0.15
-0.15
50%
25%
5%
75%
100%
Table 10-1
Reference conditions for flow calibrations (ISO 9104 and DIN/EN 29104)
Flow conditions Fully developed flow profile
Temperature,
medium
20 °C ± 2 °C (68 °F ± 3.6 °F)
Temperature,
ambient
20 °C ± 2 °C (68 °F ± 3.6 °F)
Liquid pressure 2 ± 1 bar
Density
0.997 g/cm3
Brix
40 °Brix
Supply voltage Un ± 1 %
Warming-up time 30 min.
Cable length
5 m between transmitter and sensor
Table 10-2
Additions in the event of deviations from reference conditions
Current output
As pulse output ± (0.1% of actual flow +0.05 % FSO)
Effect of ambient
temperature
● Display/frequency/pulse output: < ± 0.003%/K act. ± 3.6 °F)
● Current output: < ± 0.005 %/ K act.
Effect of supply
voltage
< 0.005 % of measuring value on 1 % alteration
Step 4: Improving the application
In the following it is described how to find the causes of a high Zero point standard deviation and
how to improve the installation.
Setting Low flow cut-off
In order to see if the zero point becomes more stable when making changes / adjustments, the
Low Mass flow cut-off (MassFlowCutOff) must be set to 0.0 kg/s.
When Low flow cut-off has been set, it is possible to see the instability directly from the mass
flow in the online window ("View → Process variables")
This information can be used to troubleshoot. For example, tightening the brackets which hold
the sensor, or turning off the pump to check if vibrations from the pump are disturbing the
sensor, etc.
Diagnostics and troubleshooting
10.3 Operation troubleshooting
SITRANS FC330
192
Operating Instructions, 01/2019, A5E44030648-AB
Содержание SITRANS FC330
Страница 14: ...Introduction 1 8 Notes on warranty SITRANS FC330 14 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 56: ...Installing mounting 4 3 Disassembly SITRANS FC330 56 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 126: ...Operating 7 2 Remote operation SITRANS FC330 126 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 158: ...Parameter assignment 8 2 Functions SITRANS FC330 158 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 216: ...Technical data 11 13 Pressure temperature ratings SITRANS FC330 216 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 224: ...Dimension drawings 12 7 Mounting bracket dimensions SITRANS FC330 224 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 236: ...HART communication B 1 Mode of operation HART function SITRANS FC330 236 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 274: ...Zero point adjustment SITRANS FC330 274 Operating Instructions 01 2019 A5E44030648 AB ...
Страница 280: ...Index SITRANS FC330 280 Operating Instructions 01 2019 A5E44030648 AB ...