Memoskop Programming
7
Standard programming
7
The SIREMOBIL Iso-C with LITHOSTAR MODULARIS is delivered with an organ program
that is programmed with one of the following default values:
The changes from Standard Organ program 1 appear in bold and are underlined.
The work procedure is described in the SIREMOBIL Iso-C user manual.
The program name is "Litho".
Organ program 1, Litho (User Setup, CTRL+U)
Mode of operation
DL
IDL
DR
SUB
Roadmap
Program name
Standard Standard Standard
Standard Standard
Dose rate
MID
HIGH
HIGH
HIGH
HIGH
SIREMATIC normal
HC2
HC2
n.a.
(DR850W)
n.a.
(Iodine)
n.a.
(Iodine)
SIREMATIC push
HC2
HC2
n.a.
(DR850W)
n.a.
n.a.
Noise red. Low
K = 4
K = 4
K = 4
n.a. n.a.
Noise red. High
K = 8
K = 8
K = 16
n.a.
n.a.
Dose reduction low
n.a.
high
n.a.
n.a.
n.a.
Dose reduction high
n.a.
high
n.a.
n.a.
n.a.
Auto LIH disk transfer
NO
NO
n.a.
n.a.
n.a.
Auto transfer each image
n.a.
NO
n.a.
n.a.
n.a.
Disk transfer rate
0
n.a.
n.a.
n.a.
n.a.
Auto disk transfer
n.a.
n.a.
NO
n.a.
n.a.
Subtraction n.a.
n.a.
n.a.
MAX
MAX
Landmark
n.a.
n.a.
n.a.
10%
NO
Video signal at document
n.a.
n.a.
n.a.
POS
POS
Image display
n.a.
n.a.
n.a.
POS
POS