Service software
3 - 13
Siemens AG
SPL1-130.840.02
Page 13 of 18
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
Software download with Windows 95 / 98 / NT
3
Do not select "ca
pture text file"
when downloading the software.
Insert the diskette containing the software into drive A
Press the "s" key and then press "1"
In the
"Übertragung"
menu
,
select
...
Datei senden
German Windows:
English Windows:
For additional procedures, refer to pages 3-2 or 3-3
Log on, as described on page 3-2 or 3-3
I
n the
"Datei senden"
window: Dateiname: click
"Durchsuchen"; select the file with the "bin" extension
on the diskette inserted (Laufwerk beachten)
Protokoll: select
Xmodem
. Click
Senden
In the
"Send File"
window
: File name: click
"Browse"; select the file with the "bin" extension on
the diskette inserted. (observe drive)
Protocol: select
Xmodem
. Click
Send
In the
"Transfer"
menu
, select ...
Send File
During the data transfer, "1" is displayed on board D3 and the decimal point flashes.
After the transfer, a test is executed. ("2" is displayed on board D3; "0" or "7" when complete.)
Disregard the error messages
Switch the system off / on
Press the "h" key to display the text
Check the system-specific settings (language, ECG triggering, pulse rate and table)
Set the date and time. Check Ultrasound default values.
No
Press the "n" key and
enter the name
Yes
Was software
transferred to a new
D3 board?
Is correct software
version displayed?
No