LITHOSTAR MODULARIS
4 - 21
Siemens AG
SPL1-130.840.02
Page 21 of 34
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
•
Depending on the language selected, the following text will appear on the control display
panel of the LITHOSTAR MODULARIS:
- "Bitte eine LITHOCARD einlegen!"
- "Please insert a LITHOCARD!"
- " Introducir una LITHOCARD!"
- "Veuiller insérer une LITHOCARD!"
•
Insert a "pay per use" card or a "gold card"
¹
If a gold card is inserted, the following - text is displayed: "New LITHOCARD
Gold" inside! Increment INIT-Counter? - Answer "Yes" to this question.
•
For Gold card only: mount the chip card reader.
•
If the ultrasound localization is available, the potentiometer will have to be adjusted
accordingly (see chapter on "Service software", select option: u).
•
Check if the default values for ultrasound are available (SPL1-130.038.01...) and
whether any necessary changes were made.
Board D3 Addition (Ultrasound)
4
If the ultrasound addition is being replaced, the potentiometer will have to be adjusted
accordingly (see chapter on "Service software", select option: u).
Pressure measurement
4
To guarantee a constant energy output, pressure measurements must be performed after
working on the shock wave system. In the case of statements of the operating staff about
deviating pressure values, the current values must be compared with the originally mea-
sured values before replacing the shock wave head.
•
Switch the system off.
•
Remove the covers of the shock wave head (refer to the corresponding section).
•
Remove the coupling bellows; when doing this, carefully pull the guide pin out of the air
suction hose (Fig. 8 and Fig. 9).
•
Loosen the four screws (V/Fig. 3) on the iso-center phantom but do not remove them.
•
Flip the iso-center phantom forward and remove it completely.
•
Mount the C-system coupling bellows for the pressure measurement.
•
Fill the coupling circuit (refer to the corresponding sections).
•
Remove any air bubbles in the coupling bellows. Install the adapter for the pressure
measurement.
•
Perform the pressure measurement according to instruction SPL1-120.074.01.
•
The instructions for performing the pressure measurement are delivered with the shock
wave pressure test device (PTU).
•
After completing the pressure measurement, restore the system to its previous status.
Coupling bellows (C-system), item number 16 12 337 must be
used in order to perform the pressure measurement.
NOTE