![Siemens MODULARIS Uro Plus Скачать руководство пользователя страница 103](http://html.mh-extra.com/html/siemens/modularis-uro-plus/modularis-uro-plus_service-instructions-manual_1266117103.webp)
MUP Menu Ultrasound Device
8 - 3
Siemens AG
SPL1-130.840.02
Page 3 of 8
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
Selection of the MODULARIS cross on the SONOLINE Prima
Ultrasound localization
8
•
In ultrasound localization a holder in which an ultrasound probe can be fastened is fitted
to the shock wave head.
Only the following probes can be used:
•
Press the “F4” key.
•
Press the "5" key.
•
Select the MUP with the trackball.
•
Press the
key.
•
Enter the password.
•
Select "Litho Mode no X-ray" with the trackball.
•
Press the
key.
•
Activate "Litho Mode no X-ray" with the trackball.
•
Press the
key.
•
Press the
key two times.
•
Deselect the screen saver on the SONOLINE Adara (refer to the SONOLINE Adara
operating instructions).
•
When a
or
probe is selected, the following cross (Fig. 3) is displayed.
Fig. 3
3.5C40S convex array 3.5 MHz
3.5/5.0 SI sector probe