MODULARIS Uro Plus
SPL1-130.840.02
Page 4 of 8
Siemens AG
Rev. 03
04.05
CS PS 24
Medical Solutions
5 - 4
Iso-center with X-Ray
•
If not, adjust the shock wave head as shown in Fig. 6 (
D or E
):
- Loosen the screws (5/Fig. 4) on both sides.
•
The shock wave head must be moved in parallel (see scale 7/Fig. 4).
The screw (6/Fig. 4) is also located on the other side of the shock wave head support.
If the shock wave head cannot adjusted, "rotate" the docking plate at the SIREMOBIL
Iso-C , i.e. the back part is higher than the front (8/Fig. 5).
•
If the connecting position of the docking plate has changed, check the settings for the lift
stop switch on the SIREMOBIL Iso-C.
- Tighten the screws (5/Fig. 4) on both sides.
Concluding work
5
•
Perform the protective conductor measurement.
•
Perform a function check.
- D:
Move the C-arm into the 20° position
as shown in (D/Fig. 6)
(viewed from the table).
- E:
Move the C-arm into the 20° position
as shown in (E/Fig. 6)
(viewed from the table).
-
Correct the position of the shock wave
head with the screw (6/Fig. 4) as
shown in (D/Fig. 6).
-
Correct the position of the shock
wave head with the screw (6/Fig. 4)
as shown in (E/Fig. 6).