MODULARIS Uro Plus
SPL1-130.840.02
Page 14 of 34
Siemens AG
Rev. 03
04.05
CS PS 24
Medical Solutions
4 - 14
LITHOSTAR MODULARIS
Charging unit
•
Switch the system off, disconnect the power plug.
•
Remove the covers (of the basic system unit), disconnect the charging unit power cable.
•
Wait at least 10 minutes after switching the system of. Then remove the cover on
the back high-voltage cable connection (Fig. 13).
•
Remove the high voltage cable (HK/Fig. 13) and the protective conductor (SL/Fig. 13) by
removing the screws.
•
Remove the fiber optic cable from the charging unit.
•
Remove the six Allen screws (IS/Fig. 14) (three on each side, right and left).
•
Remove the old charging unit from the basic system unit and install the new one.
•
Reattach the six Allen screws (IS/Fig. 14) and tighten them (three on each side, right and
left).
•
Reconnect the fiber optic cable to the charging unit (check for correct connections).
•
Reattach the protective conductor and the high-voltage cable with the screws (Fig. 13).
•
Reattach the cover on the high-voltage cable connection.
•
Reconnect the power cable of the charging unit.
•
Record the installation date on the charging unit (next to the capacitor).
•
Switch the system on, read out the counters for the charging unit and the spark gap and
record the values on the
old
charging unit. Reset the counter (refer to chapter 3).
•
Perform a function test.
•
Switch the system off and reattach the system cover.
Fig. 13
Fig. 14
The charging unit should not be opened again.
If a charging unit with spark gap is replaced by a charging unit
with semiconductor switch (11 58 000), then the software VA00J
must be used.
IS
SL
HK
N
NOTICE