Iso-center with ultrasound
6 - 7
Siemens AG
SPL1-130.840.02
Page 7 of 16
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
Checking the target on the Sonoline Adara
•
Set the value (D/Fig. 10, 11) to 180 using the knob
.
•
Slide the ultrasound probe all the way forward to the shock wave head.
•
Press the
key and select the measuring point
.
•
Use the track ball to move to the center of the crosshairs. In the display (F/Fig. 10), a
value of - see Table - must appear that is relative to the probe.
•
Press the
key ("Freeze" is deselected).
•
Move the ultrasound probe all the way in the other direction.
•
Press the
key and select the measuring point
.
•
Use the track ball to move to the center of the crosshairs. In the display (F/Fig. 11), a
value of - see Table - must appear that is relative to the probe. If this is not the case, the
adjustment procedure must be repeated for the potentiometer in the ultrasound arm or
the ultrasound arm must be replaced.
•
Press the
key ("Freeze" is deselected).
Fig. 10
Fig. 11
The isocenter must be checked with every probe type and after
every replacement of a probe.
Probe
Changeover with
Value for D1
Sector (1/Fig. 2)
15 ± 1 mm
with default values,
see SPL1-130.038.01
Array (2/Fig. 2)
13 ± 1 mm
Probe
Changeover with
Value for D1
Sector (1/Fig. 2)
153 ± 1 mm
with default values,
see SPL1-130.038.01
Array (2/Fig. 2)
151 ± 1 mm