MODULARIS Uro Plus
SPL1-130.840.02
Page 2 of 16
Siemens AG
Rev. 03
04.05
CS PS 24
Medical Solutions
6 - 2
Iso-center with ultrasound
Checking the target on the Sonoline G20
6
•
Set the value (1/Fig. 3, 4) to 180 with the "Image field/zoom" knob.
•
Push the ultrasound probe as far as possible up to the shock wave head.
•
Press the "Measurement" key and select the "Set" measuring mark.
•
Move to the center of the crosshair with the trackball. A value of 13 mm + 1 mm (2/Fig. 3)
must appear in the display (2/Fig. 3).
•
Press the "Measurement" key ("Freeze" is deselected).
•
Shift the ultrasound probe as far as possible in the other direction.
•
Press the "Measurement" key and select the "Set" measuring mark.
•
Move to the center of the crosshair with the trackball. A value of 151 mm + 1 mm must
appear in the display (2/Fig. 4).
•
If this is not the case, the adjustment process for the potentiometer in the ultrasound arm
must be repeated or the ultrasound arm must be replaced.
Fig. 3
Fig. 4
The isocenter must be checked with evry type of probe and after
every probe change.
The designations of the keys can vary according to language
version.
Comparison table, see SONOLINE operating instructions.
1
2
1
2
NOTE
NOTE