MODULARIS Uro Plus
SPL1-130.840.02
Page 12 of 18
Siemens AG
Rev. 03
04.05
CS PS 24
Medical Solutions
3 - 12
Service software
Every probe can be set to one of three settings:
By selecting the keys "t, l, h " (depending on application) the value is reduced; select T, L,
H to increase the value.
select option: r
3
•
Press the
"r" key
to change the "pulse rate"
Selection: fixed 2 Hz (USA) or variable 1/1.5 /2 Hz
After pressing the "h" key, display for "selection option r" shows the pulse rate set.
select option: s
3
Software downloads are possible only if the software and hardware are compatible.
•
Press the
"s" key
to display the following text:
After "1" has been entered, the software download must start within 60 seconds.
If the 60 seconds elapses without starting the download, reselect "h" from the entry menu,
reselect "s", etc.
Once the software download is complete, check that the values in the "ultrasound localiza-
tion" menu for the applicators are identical to those in document SPL1-130.038.01..
(Intranet). If this is not the case, make the appropriate changes.
threshold value
fig. 2
The point at which the "scale high" value is used (default 100 mm).
scale low
fig. 2
Correction factor (default 7.7 %) between 0 mm and the "threshold
value " (This value is useful for thin patients).
scale high
fig. 2
Correction factor (default 7.7 %) between "threshold value " and end
(This value is useful for obese patients.).
Software download.
If not observed, D3 board can be damaged.
The system and service PC must not be turned off during a soft-
ware download. If this occurs, board D3 must be replaced.
CONVEX APPLICATOR
-----------------------------------------
actual values change with key (-) (+)
-----------------------------------------
threshold value = 100 mm t T
scale low = 7.7 % l L
scale high = 7.7 % h H
CAUTION
=> S
press key '1' to confirm download, any other key to abort