
94
Final procedures
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 94 of 116
Medical Solutions
Checking the protective grounding resistance
0
Using the protective ground wire tester, measure the resistance between the protective
ground terminal in the stand (Mains input terminal) and all metallic parts of the equipment
which the patient/operator may come in contact with, incl. all object tables.
The protective ground resistance must not exceed
0.1 Ohm