Image Quality
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 86 of 116
Medical Solutions
These settings are the default ones, available upon installation
L
oading the sample images
If the sample images are not loaded by the system, proceed as follows.
Set Direct
Radiation
Black
ON
ON
ON
OFF
ON
OFF
OFF
OFF
Window
Mode
ON
ON
ON
OFF
ON
OFF
OFF
ON
Edge
Enhance-
ment Gain
0.3
0.3
0.3
0.3
0.3
0.1
0.0
0.3
Edge
Enhance-
ment Kernel
15
15
15
15
15
15
15
15
Edge
Enhance-
ment LUT
index
0
0
0
0
0
0
0
0
LUT index
18
18
18
18
20
6
6
6
Harmoniza-
tion Amplifi-
cation Gain
0.7
0.7
0.7
0.7
0.6
0.2
0
0.6
Harmoniza-
tion Kernel
Size
127
127
127
127
127
127
127
127
Factor make
window
width wider
1.6
1.4
1.8
2.0
1.5
2.0
2.0
1.6
Window
Center Shift
150
120
180
-50
140
0
0
150
Implant Rec-
ognition
Mode
ON
ON
ON
ON
ON
OFF
ON
ON
Implant Rec-
ognition Sen-
sitivity Factor
0.0001
0.0001
0.0001 0.0002 0.00025
0.0001 0.0001 0.0003
meanfac_dir_
rad_prior_lut
0.5
0.5
0.5
0.35
0.5
0.35
0.35
0.35
Parameter
Normal
High
Low
Mag/S
pot
Implant
Speci-
men
Phan-
tom
Safety