Vertical adjustment
CAUTION
Remove the protective strips before performing the vertical adjust-
ment of the swivel arm.
If not removed, the protective strips could be damaged.
¹
Make sure the protective strips are removed before
adjustments, see "Metal Curtain" on
1. Press the buttons for upward and downward movement (
) of the swiv-
el arm system.
The movement should stop automatically when either the end position is reached or
when the button is released.
NOTE
All eight vertical adjustment buttons (four on either side
) must be checked.
The switching off in upper and lower end position is effected by
switches S881 and S884 respectively.
Emergency stop
1. Press the emergency stop button
(4/Fig. 19 / p. 37)
.
The button should latch and all motorized movements of the equipment should be
blocked ("Er 813" should appear). To reset the emergency stop button, turn it
clockwise.