58
Checking dose settings
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 58 of 116
Medical Solutions
Checking the AEC dose settings
0
The dose values for the AEC settings (
H & D
selection on the control console) have to be
checked.
As a default from the factory, the following
BRICK / AEC Configuration
values are set
(for further details see the
Software Brick V...
manual SPB7-250.816.04...):
•
H = (80, 70, 45)
•
D = (100, 80, 60)
Checking the H & D dose settings for Mo/Mo
.
Proceed as follows:
1. Mount the collimator mounted plexi phantom and the compression plate simulator.
2. Register or create a patient and submit with
Exam
.
Fig. 38: Calibration_Patient register