64
Image Quality
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 64 of 116
Medical Solutions
2. Select the Procedure
VendorProc 1
from the examination tab card.
Fig. 42: Select procedure for grid line test
NOTE
Remember you can always add or remove images to the “Proce-
dure” by right-clicking the mouse.
3. Select the exposure parameters:
28 kV
and
Mo/Mo
.
4. Make sure that a 1 : 1 view is selected to have a better view.
Select
View -> a 1:1
5. Make all exposures according to Tab. 1.
6. Activate magnification
(Image -> Magnify by 2.0)
and look for grid lines over the plexi
phantom area.
7. Make a not of the results (visible grid lines) in Tab. 1; choose the next image in the list
and start over with step 5 until all images are evaluated.
Test 1 protocol
0
Tab. 14
Checking the visibility of grid lines
Exposure
Exp Mode
mAs
PMMA /
mm
Tube Angle
Visible grid lines
(yes/no)
1
mAs
32
30
0
o
2
mAs
56
40
0
o
3
mAs
110
50
0
o