34
Mechanical covers
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 34 of 116
Medical Solutions
CAUTION
The front cover must not press against the collimator.
Risk of damage.
¹
Be careful while attaching the front cover.
4. Fasten the front cover to the side covers with two short screws (see
).
Stand covers
0
WARNING
The edges of the metal curtain on the stand are very sharp.
They may cause severe injury.
¹
Remove the protective strips carefully when the covers
are to be mounted. Store the strips in the holders pro-
vided on both sides of the curtain.
NOTE
When mounting the stand covers, make sure that the screws at the
top of the stand are securely tightened and have functioning con-
tact washers.
These screws are used to establish protective ground connec-
tions.
Fig. 16: Mounting the stand covers
1. The covers are delivered with a protective plastic film, which should be removed and
discarded.
2. Make sure that clips are positioned over the holes where the screws (see
) are to be fastened. These clips enable the screws to fasten the cov-
ers.
3. Fit left and right front covers (see
) to the stand and fasten with screws
and contact washers (see
).