
Siemens AG
SPB7-250.815.06.02.02
MAMMOMAT Novation DR
12.05
CS SD 24
Image Quality
71
Page 71 of 116
Medical Solutions
•
Connect the oscilloscope as follows:
Channel 1 to measuring point HV_ACT on the D750, 1 V/T.
Channel 2 to measuring point Grid_M on the D802, 10 V/T.
Select time base 0.4 s/T.
•
Switch the System
ON
•
Select 28 kV, Mo/Mo, 500 mAS; release an exposure
•
Evaluate the oscilloscope diagram.
The period of the high voltage in a steady state condition (after the 3rd period) has to be
between 500 ms and 750 ms. See diagram:
Fig. 50:
Grid speed
If the grid lines are still visible after this process , increase. i.e. ( change) the grid slow
speed from 40% to 50% and run this test again.
Checking Image Quality
0
While calibration is typically performed by the customer as part of normal system opera-
tion, calibration information is provided in the event that you are required to perform a sys-
tem calibration.
To calibrate the system, proceed as follows:
•
First, perform a
gain calibration
of the detector and check the images for bad pixels.
•
If you observe any bad pixels in the image, create the
bad pixel mapping
and perform
a
gain calibration
of the detector afterwards.
If you have to troubleshoot image quality problems, please see the chapter "Trouble-
shooting image quality" on
(Troubleshooting image quality / p. 108)
.