28
Detector installation
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 28 of 116
Medical Solutions
Mounting the grid
0
The materials needed for the grid installation can be found in the accessory material box.
Fig. 12: Grid installation
1. Insert the left and right grid slider
from the back into the rail.
2. Insert the left and right grid holder into the grid slider
.
3. Screw the grid
to the grid holder
. The marking has
to be on top, text must be readable.
4. Line up the grid so that it runs parallel to the front of the table; the front switch has to be
engaged and the grid sliders have to be engaged with the spring
.
NOTE
Make sure that the grid sliders are locked in place and have only
little play.
Make sure the springs move up and down easily.
5. Mount and dismount the MAG table several times and check if both grid holders
are engaged with the spring
.
NOTE
The magnification table is optional. When the magnification is not
supplied with the system, make sure that the grid is locked into the
front of the table by the spring
.
6. Connect the
cables for the magnification table on both sides accord-
ing to the labels.