78
Image Quality
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 78 of 116
Medical Solutions
5. Select
Patient -> Calibration
from the syngo menu bar.
Fig. 57: Calibration
6. Click on the
Bad pixel
button to start the bad pixel mapping.
7. Confirm the message that pops up.