Power up the MAMMOMAT
17
Page 17 of 116
Medical Solutions
4-
3Power up the MAMMOMAT
Radiation protection door switch (optional)
0
It is optional to connect a radiation protection door switch to the MAMMOMAT Novation
DR
master board D750 connector X754.
NOTE
A radiation protection door switch is not delivered with the MAM-
MOMAT Novation
DR
.
A 15V / 5 mA signal is used in the door switch circuit. When the signal path is opened by
the door switch, the exposure will immediately be terminated.
Fig. 3:
Connecting the door switch
1. To connect a door switch, remove the jumper from the X754 (Pin 1,2) located on the
master board D750, see Fig. 1, and connect the door switch instead.
If a radiation protection door switch is not used, a jumper, see
, on the mas-
ter board 750 at connector X754 (Pin 1,2) has to be inserted.