MAMMOMAT Novation DR
12.05
CS SD 24
Appendix
109
Page 109 of 116
Medical Solutions
Line Artifacts
Problem
Cause
Recommended Action
Line Parallel to the
Short Axis
Seam:
•
Seam correction algorithm not
turned on; wrong algorithm se-
lected.
•
Detector gain not calibrated
properly.
•
Seam (lines 1535 and 1536)
needs to be mapped out.
Source lines:
•
detector not calibrated properly.
•
line needs to be mapped out.
•
Perform detector calibration pro-
cedure (refer to "Gain calibration"
on
.
•
Perform detector calibration pro-
cedure (refer to "Gain calibration"
on
).
•
Perform the bad pixel correction. (
Refer to "Creation of new pixel
map" on
).
Line Parallel to the
Long Axis
Line ends at seam or line partially
defective (gate lines).Line ends at
seam or line partially defective (gate
lines).
•
Detector not calibrated properly.
•
line needs to be mapped out.
•
Perform detector calibration pro-
cedure (refer to "Gain calibration"
on
).
•
Perform the bad pixel correction. (
Refer to "Creation of new pixel
map" on
).
Line Runs Through
Seam
•
Read out problem; HV setting
time too short.
•
Hard copy device or processor
artifact.
•
Follow applicable problem solving.