Image Quality
79
Page 79 of 116
Medical Solutions
8. As a result, the right screen will change as shown in the following figure.
Fig. 58:
Accept_reject screen
9. Set the following exposure parameter on the control console of the MAMMOMAT Nova-
tion
DR
Mo/Mo, 28KV, 250mAs
10. Insert the collimator mounted plexi phantom.
11. Mount the compression plate simulator.
12. Double-click the
first
Image of
Procedure: Calibration
.
13. Start an exposure.
Wait until the image is loaded into the viewer. The following lines are displayed in the
message bar.
Image acquisition is in progress .....
Saving processes Image .....
Loading image .....
14. Click the
Accept
or
Reject
button.
If you
accept
the image, the new pixel map is created. When the
Accept images
counter increases to
1
, no further images need to be taken.
If you
reject
the image, the
Bad Pixel Mapping
menu is displayed. Correct the problem
and return to step 13.
NOTE
It is recommended to reject an image if it shows any cut-off edges
due to collimation or misalignment, or if there are any artifacts
from debris or obstructions.
15. Select
Examination -> End examination
from the syngo menu bar to exit the Calibra-
tion.