Final procedures
WARNING!
When attaching the front cover, make sure that all eight contact washers (four on either
side) are used again as in the original assembly (under the screws with clearance holes
in the plastic strips). The screws must be securely tightened and the washers must be func-
tional. This is necessary to establish protective ground connection.
2
Press on the outer cover strip (1).
Final protective grounding resistance test
Using the protective ground wire tester, measure the resistance between the mains
grounding connection (incl. mains cable) and any metallic part of the stand and generator
which the patient/operator may come in contact with.
The protective ground resistance must not exceed 0.2
Ω
.