Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
Preparatory work
3-5
CAUTION!
The protective strips must be removed from the metal curtain before any vertical adjust-
ment of the swivel-arm system is performed.
Holders (2) for storing the protective strips when not in use, are provided on both sides of
the curtain, see Figure 5.
Unpacking the generator and mounting the radiation
shield
Figure 6 Unpacking the generator
1
Open the cardboard box by cutting the plastic straps.
2
Lift out the lead-glass pane and put it away in a safe place.
3
Remove the protective cover from the generator.
4
Upend the generator (1) and lift it off the pallet (2) (two persons are required).
5
Mount one stanchion (3) for the lead-glass pane (4) onto the generator.
MAM00085
3
4
3
2
1