Final procedures
Removing the measuring instruments
1
Turn OFF the generator.
2
Remove the service PC.
3
Remove connected measuring instruments.
Checking the protective grounding resistance
Using the protective ground-wire tester, measure the resistance between the protective
ground terminal in the stand (X899, see Section 5) and all metallic parts of the equipment
which the patient/operator may come in contact with, incl. all object tables. The protective
ground resistance must not exceed 0.1
Ω.
Mounting the cable duct covers
If it´s not already done mount the cable duct covers, see section 4.