Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
Cable connections
5-9
Figure 9 Incoming mains 2-phase
Connecting the mains supply to the generator (400 V, 2-phase)
Figure 10 Mains supply to the generator
1
Check that there is no wire jumper at F20 (1) and no brass plug at F5 (2).
Fuses must be inserted into fuse holders F20 and F5.
2
Connect the mains supply cables between stand and generator to fuses F10 and F20
on the generator.
Wire marked L1 to F10 and L2/N to F20.
3
Connect the protective ground wire from the stand to the generator (3).
TWO-PHASE
GENERATOR
STAND
F10
F20
X881
Incoming mains
208, 230, 240, 277 or 400 VAC
P.E N
L1 L2 L3
X899
MAM00095
MAM00103
2
F10
F20
1
F1
F2
F3
F4
F5
3