Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
Further programming
13-1
Setting the real time clock
Prerequisite
The correct date and time must be set in the service PC. If not set date at C:\> date and
time at C:\> time.
Setting the time in the mammomat
Main menu:
1
Select Configuration
<↑> <↓>
< ENTER>.
2
Select Clock
<↑> <↓>
< ENTER>.
3
Press <F2>.
With <F2> the date and time set in the service PC are taken over by the Mammomat.
Setting the tube-current reduction
In Mammomat 300, the tube-current reduction is factory-set to "ON", which means that
the tube current is reduced to approx. 50 mA during the first 100 ms of the exposure. In
this way, possible grid lines are avoided.
If the tube-current reduction is set to "OFF", grid lines may occur at short exposure times
(< 100 ms).
NOTE!
The tube-current reduction is effective only in Iontomat mode.
Main menu
1
Select Configuration
<↑> <↓>
< ENTER>.
2
Select Miscellaneous
<↑> <↓>
< ENTER>.
3
Select Tube mA red.
<↑> <↓>
< ENTER>.
Tube-current reduction (TCR) at short exposure times.
TCR switch is ON/OFF.
Exposure time with 50 mA tube current is 100 ms.
4
Switch ON or OFF with
<↓>.
5
Save the programmed values with <F2>.