65 02 186
Mains connection and power supply
7-3
7
Mains supply "ON" (Do not switch the system on yet).
8
Measure the voltage at transformer T1 (6) terminals 1 and 3, transformer T1 is located
under the line filter Z1 (5). There must be 195 V- 253 V between the terminals.
9
Measure the voltage at the output of line filter Z1 it must lie within 195 V- 264 V.
Checking the supply voltages
The correct supply vo5 V, +15 V and +24 V have already been tested in the fac-
tory. Only a visual check of the function is therefore needed.
1
System "ON".
2
Check that the following orange diodes are "ON" on p.c. board D704:
•
V39 (+15 V)
•
V38 (-15 V)
•
V41 (+5 V)
•