Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
Field light
15-1
Checking and adjusting the field light time
Checking the field light time
Figure 1 Tube head
1
Switch on the field light by pressing switch (1) or by pressing the compression foot
switch (2).
The field light shall go out automatically after a preset period of time (normally 20
sec.).
Figure 2 Foot switch
MAM00141
1
MAM00133
2