Start-up and functional test of the IONTOMAT
12-9
Checking the automatic transparency adaption
Check the automatic transparency adaption with a grid table with 20, 40 and 60 mm plexi
at the voltage(s) preferred by the customer (within 25-35 kV). If the dose rate is too low
(error 013 or "Limit"), use 50 mm plexi instead of 60 mm.
Tolerance:
≤
0.3 density difference at constant voltage and thickness 20-60 mm.
≤
0.3 at constant thickness and 25-35 kV.
NOTE!
Tolerance valid for grid tables and film/screen combination according to table 1.
These limits apply for film with max. gamma 4.0 measured at the density chosen.