Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
Start-up and functional test of the IONTOMAT
12-1
Checking and programming with the service PC
Use the service PC to check and program the IONTOMAT. Operation of the service PC
is described in section 21.
The tables are optimized for the following four film/screen combinations. Curve x1 is
used for magnification tables and curve x2 for other object tables (with or without grid).
Table 1
Program the IONTOMAT values as follows:
Sensitivity
Start with the sensitvity listed in the table for channel B, both H (
) and D
(
), see also M300 Service Instructions. If the film/screen combination used by the
customer is not listed in table 1, start with sensitivity 11.
NOTE!
Be sure to use the same type of cassette, screen and film in each object table as the
customer uses.
Indicates the sensitivity for the IONTOMAT channel for contact exposure tables
(H) and magnification tables
(D).
Values from 0 to 31 can be entered.
Channel A is not used, the sensitivity is therefore set 0.
Dupont Microvision C
Agfa Mamoray MR6
Fuji UM-MA HC
Kodak Min-RE
Correction curve
Sensi-
tivity
Magni-
fication
Non-mag-
nification
Film
Screen
Cassette
Kodak Min-R
Kodak Min-R2
11
A1
A2
Fuji HR Mammo Fine
Fuji EC-MA
9
B1
B2
Agfa MR Detail S
Agfa MR Detail S
6.5
C1
C2
Dupont Quantavision ME Dupont Quantavision 12
D1
D2
MAM00129