Installing the generator and the stand
4-1
Arranging the components
1
Position the stand and the generator in accordance with installation plan and Figure
1 below.
Figure 1 Installation, top view
Cable duct (1) and generator (3) with radiation shield (2) can be installed optionally
to the left or to the right of the the stand. Maximum cable length between stand and
generator is approximately 3.5 meter.
Radiation shield (2) perpendicular to visual line between generator and focus of the
tube head (4) according to Figure 1.
Recommended minimum distance to wall and generator base plate respectively is 700
mm (shorter distances may be used at the customer´s desire, if in compliance with lo-
cal regulations).
For the installation of the cable duct, a minimum distance of 150 mm between wall
and stands is required.
640
1400
700
700
700
700
540
1091
150
708
1
2
3
MAM00088
4
90