Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
16-2
Checking the compression device and OPCOMP
6
Press the foot switch for motorized decompression to move the compression plate up-
wards while at the same time pressing the compression plate in the opposite direction
by hand.
The upward movement of the compression plate shall now stop (safety switch).
7
Set the compression to 7 kg by using the presetting knob (5) in Figure 1.
The preset compression force appears on the display at the lower part of the stand(1)
in Figure 3.
Figure 3 Display at the lower part of the stand
NOTE!
The display (1) switches over to the actual compression-force value, when the compres-
sion plate comes in contact with the breast. The compressed thickness is shown on the dis-
play (2).
8
Press the foot switch for motorized compression to move the compression plate
downwards while simultaneously pressing the compression plate in the opposite di-
rection by hand.
The compression plate shall now slow down and stop, when the preset compression
force is reached.
NOTE!
During compression, the motorized vertical adjustment as well as the rotary motion of the
swivel-arm system shall be blocked.
F
mm
o
MAM00134
OC
1
2