Siemens-Elema AB
SPS-UD
RX B7-120.033.01.10.02
MAMMOMAT 300
Solna, Sweden
05.98
65 02 186
12-2
Start-up and functional test of the IONTOMAT
Sensitivity correction
Program "Sensitivity correction" to 0 for "without grid" and 0 for "magnification".
Indicates the sensitivity correction for different types of object tables.
The position "without grid" applies to tables with as well as without grid.
Factory set "0", "0", "0".
Different film-density corrections can be programmed for the different tables with the
"Sens-cor" module.
Correction curve
Program "Correction curve" according to table 1 above (see also M300 Service Instruc-
tions). If the customer uses another combination or mixture of those listed in tale 1, please
refer to page 12-3.
Indicates which correction curve shall be used for contact exposure tables
(H)
and magnification tables
(D). Values shown as delivered.
MAM00127
IONTOMAT PM - Sens. correction
Without grid
1 Help
2 Save
3
4
5
6
7
8
9
10 Quit
With grid
Valid entries from -8 to 8 in 1/2 E.P.
For pos entries 1 st digit is < space >, for neg entries type < - >.
Input of decimal point not required!
Magnification
24 x 30
_.0
.0
.0
MAM00128
IONTOMAT PM - Correction curve
Curve number
1 Help
2 Save
3
4
5
6
7
8
9
10 Quit
Speed H
Speed D
A1
Press space to change the correction-curve table.
_A2