Annex
78
© SICK AG · Division Auto Ident · Germany · Subject to change without notice
8013796/UL20/2010-09-27
Operating Instructions
Laser Measurement Systems of the LMS500 Product Family
Chapter
10
10.2.2
Stop measurement
The LMS stops measuring.
Important
The required user level is “Authorised client” (see
section 10.2.7 “Select user level/login”
Request
Message structure:
sMN LMCstopmeas
Answer
Message structure:
sAN LMCstopmeas
ErrorCode
Example
Request:
<STX>sMN{SPC}LMCstopmeas<ETX>
HEX string:
02 73 4D 4E 20 4C 4D 43 73 74 6F 70 6D 65 61 73 03
Answer:
<STX>sAN{SPC}LMCstopmeas{SPC}0<ETX>
HEX string:
02 73 41 4E 20 4C 4D 43 73 74 6F 70 6D 65 61 73 20 30 03
Message part
Description
Variable type
Length
(byte)
Value range
Type of command
Request (SOPAS method by name)
string
3
sMN
Command
Stop measurement
string
11
LMCstopmeas
Message syntax III:
Request “Stop measurement”
Message part
Description
Variable type
Length
(byte)
Value range
Type of command
Acknowledgement of receipt
(SOPAS answer)
string
3
sAN
Command
Stop measurement
string
11
LMCstopmeas
ErrorCode
The command has been accepted
if the error code 0 is returned.
Enum8
1
0
no error
1
error, status change not
permitted
Message syntax IV:
Answer to the “Stop measurement” request