Manually set the desired distance for “4 mA”
Operating example or steps:
RUN mode
→ Q
1
Output
/
to → Q
2
Output
→ Function Q
2
/Q
a
/
to → Manual Q
a
→
0
0200 (Note:
0
= flashing 0)
/
,
/
,
/
,
/
,
/
→ desired distance in mm
→ Saved
to → RUN mode
Necessary default settings:
Factory setting or function Q
2
/Q
a
=
4-20 mA
Mode Q
a
= Distance (only for user
level = Advanced)
Factory setting:
200 mm
Manually set the desired distance for “20 mA”
Operating example or steps:
RUN mode
→ Q
1
Output
/
to → Q
2
Output
→ Function Q
2
/Q
a
/
to → Manual Q
a
→ 4 mA
/
to → 20 mA
→
3
0000 (Note:
3
= flashing 3)
/
,
/
,
/
,
/
,
/
→ desired distance in mm
→ Saved
to → RUN mode
Necessary default settings:
Factory setting or function Q
2
/Q
a
=
4-20 mA
Mode Q
a
= Distance (only for user
level = Advanced)
Factory setting:
30,000 mm
Activate the “0-10 V” output function or analog voltage output
Operating example or steps:
RUN mode
→ Q
1
Output
/
to → Q
2
Output
→ Function Q
2
/Q
a
→ 4-20 mA
/
to → 0-10 V
→ Saved
to → RUN mode
Necessary default settings:
-
Factory setting:
4-20 mA