53
14. Application Functions
Read Chapter 7 Before Use of
These Pages
14. Application Functions
Application measurement functions are described in this chapter.
Only one of the functions in the menu group 4 (Refer to
) (
to
) can be used at a time. When
one of the functions in menu group 4 is to be used with a weighing unit other than gram, select the
function from the gram-display first. Then, change to the other unit with the
[UNIT]
key. To release this
function, use the
[POWER]
key. For returning to the previous function, Last Menu Recall function is
convenient. There are some other combinations of functions that can be used together. Some func-
tions employ the value set in Zero Range setting (refer to
14.1 Solid Specific Gravity Measurement
Solid specific gravity measurement refers to the measurement of the sample (solid) weight in the air
and in a liquid of known specific gravity (or density) and the calculation of the sample specific gravity
(or density).
The
T
d symbol is used to represent the solid specific gravity in this balance. The data output unit is DS.
1
Select menu item
in advance.
2
Remove the below-weigh hook cap from the
bottom of the balance to expose the below-weigh
hook.
Note
Use of the optional
SMK-101, or SMK-102 Specific Gravity Measure-
ment Kit
(refer to
) is recommended for efficient measurements.
When using the SMK-101 or SMK-102, refer to the instruction manual of
the kit. Follow the instruction below when a hanging pan and a tank are
prepared by the user.
41
53
60
Note
Enter the value of the specific gravity
(
or density (g/cm
3
)
)
of the liquid
(water, alcohol etc.) in which the sample is immersed. Refer to
for
numerical value input. To cancel the set-up, set the value to zero.