58
14. Application Functions
Read Chapter 7 Before Use of
These Pages
14.4 Interval Timer
Automatically outputs the displayed value at preset intervals. The “T” symbol (“T” of the Tare symbol) is
illuminated when the Interval Timer is activated.
The optional AKB-301 Application Keyboard is convenient when the interval is changed frequently.
1
Select menu item
and set the output inter-
val (00:01 = 1 sec to 99:59 = 99 minutes 59 sec-
onds).
2
In the interval timer standby state when the “T”
and the Stand-by symbols are both illuminated,
press the
[PRINT]
key.
The first data is output. Data will be automatically
output at the set time intervals thereafter.
3
To stop output, press the
[POWER]
key.
The balance returns to the interval timer standby
state in step 1).
50
Notes
• Use the
[O/T]
key to erase the tare or zero the balance at any time.
• Pressing the
[POWER]
key in the interval timer standby state brings the
power supply standby state.
• To release the interval timer function, keep pressing the
[POWER]
key.
This does not reset the interval timer function to zero.
• Using the interval timer function to record data over a long period may
cause data error due to balance drift.
• Some instruments receiving the data may not operate normally if the set
time interval is short. To correct this, set the time interval to a longer
period. When the set time interval is short and the instrument connected
to DATA I/O is unknown, it is recommended to set the handshake to a
setting other than “H-tm”
.
76