32
10. Calibration
Read Chapter 7 Before Use of
These Pages
10.2 Calibration Execution
10.2.1
Span Calibration Using the Built-in Weight (UW Series Only)
1
Verify that the balance is in mass display and
that the pan is empty.
2
Press the
[CAL]
key once. “i-CAL” is displayed.
(If “i-CAL” is not displayed, return to mass display
and select menu item
.)
3
Press the
[O/T]
key.
After “i-CAL3”, “i-CAL2”, “i-CAL1”, “Set”, “CALEnd”
the mass display will appear indicating the com-
pletion of span calibration.
Notes
• Setting before shipment is as the following:
UW series: Span calibration using the built-in weight
UX series: Span calibration using external weights
The type of calibration can be changed (See
).
• Calibration will not be performed when the weight on the pan is not near
zero, or the balance is not stable.
...
1