10. Calibration
Read Chapter 7 Before Use of
These Pages
10.3.5 PCAL Password Setting (UW series only)
This password is necessary to access the PCAL function.
It is recommended that the balance administrator set this password to prevent an unauthorized person
from incorrectly calibrating the built-in calibration weight.
Select menu item
. The numerical setting display appears.
Enter a 4-digit number from “0000” to “9999”.
Refer to
Notes
• “SEt” is displayed during the process. Leave the balance in a stable state
until the mass display appears as in step 5.
• In PCAL, the value of the “weight to be loaded” cannot be changed.
• Set the PCAL password using menu item
.
Note
When the menu is reset, the PCAL password is reset to “9999”.