57
14. Application Functions
Read Chapter 7 Before Use of
These Pages
14.3 Peak Hold
Measures the displayed peak value. The “P” symbol ( “P” of Auto-Print symbol) is illuminated when the
Peak Hold function is activated. “Peak value” is the highest or lowest stable value displayed after the
display has changed beyond five times the Zero Range.
Select menu item
in advance.
1
In the peak detection standby state with the “P”
symbol and the Stand-by symbols illuminated,
press the
[O/T]
key to tare the display.
2
Press the
[PRINT]
key.
The Stand-by symbol disappears and peak value
detection starts.
3
The “P” symbol and Asterisk are simulta-
neously displayed after the peak value is
detected, and the data is output. This display will
not change regardless of the load on the pan.
4
Press the
[POWER]
key.
The balance returns to the peak detection
standby state in step 1).
49
Notes
• Press the
[POWER]
key in the peak detection standby state to initiate
the power standby state.
• Press the
[POWER]
key during detection of the peak to return to the
peak detection standby state.
• Polarity of the peak value displayed is “polarity of the displayed value of
the first change by five times or more of Zero Range from the display
within Zero Range.”
• Usually the peak value is easily measured by selecting menu item
(Pouring mode). Depending on the weighting conditions and the sample
type, this setting may not always be suitable.
• Pretaring Value (13.7) function cannot be used with Peak Hold function.
23