14. Application Functions
Read Chapter 7 Before Use of
These Pages
3
Hook the hanging pan, and then immerse the
hanging pan in the tank filled with the liquid of
known specific gravity (or density).
4
From mass display, press the
[UNIT]
key sev-
eral times until
T
d (inverse triangle and “d”) is
displayed.
5
Press the
[O/T]
key.
6
Load the sample on the balance pan (or in the
hanging pan in air).
7
After the stability mark illuminates, press the
[CAL]
key.
“dSP oL” may be displayed but this does not indi-
cate a malfunction.
8
Load the sample on the hanging pan immersed
in the liquid. The specific gravity (or density) of the
sample is displayed.
9
Repeat steps 5 through 8 for each additional
sample.
Notes
• Up to four decimal places are displayed for specific gravity. When it is
not possible to stabilize the balance in all 4 decimal places, use the 1d/
10d switching function.
• When loading the sample on the pan in the liquid, ensure that the entire
sample is immersed in the liquid.
• The balance does not re-zero when the
[O/T]
key is pressed in this func-
tion.
• The dimensions of the below-weigh hook are shown in A-7.