13. Enhancing Productivity
Read Chapter 7 Before Use of
These Pages
13. Enhancing Productivity
Functions that are useful mainly in production sites are described in this chapter.
Only one of the functions in the menu group 4 (Refer to
) (
to
) can be used at a time. When
one of the functions in menu group 4 is to be used with a weighing unit other than gram, select the
function from the gram-display first. Then, change to the other unit with the
[UNIT]
key. To release this
function, use the
[POWER]
key. For returning to the previous function, Last Menu Recall function is
convenient. There are some other combinations of functions that can be used together. Some func-
tions employ the value set in Zero Range setting (refer to
13.1 Checkweighing and Target Display
The UW/UX series balance has an analog bar graph located on the left side of the display. This graph
can be conveniently used for checkweighing or cumulative weighing.
The graphic
display functions include the two display modes of checkweighing, the target mode, and
full scale mode. Only one of them can be used at a time. Refer to
for the full scale mode and the no
bar graph display.
Upon selection of either the checkweighing mode or the target mode, the numerical value for threshold
or target is requested. Numeric values are memorized independently for each graphic display mode.
When one of the checkweighing modes is selected, a signal corresponding to HI, GO, or LO in the-
graphic display can be transmitted via the RS-232C/AUX connector.
Target, limit, upper, and lower values are set as numeric values only. Set the correct numerical value
for the unit that will be used for weighing.
• In weighing mode, changing the displayed unit does NOT change the target, limit, upper, or lower
numeric values. For example, when the upper limit has been set at 10g, switching the unit from “g” to
“kg” does not change the limit to 10kg. It will be 0.01kg.
The decimal point is invisible in the value setting display. Determine its position based on the resolution
of the balance. For example, if the balance’s minimum display is 0.01g, 150g must be set by inputting
“15000” in the display.