13. Enhancing Productivity
Read Chapter 7 Before Use of
These Pages
Print on “GO” judge of checkweighing
: Select menu item
.
When the graphic display is in one of the checkweighing modes and the stability symbol
lights in the GO range, the data is output once.
The next data output is executed after the displayed value is within the Zero Range.
For automatic outputs at preset intervals, see
Interval Timer.
13.4 Auto Zero
When the displayed value is within the Zero Range and the stability mark has been illuminated, zeroing
occurs automatically. The Zero symbol appears in the display when the Auto Zero function is active.
Other keys function as expected with the Auto Zero function activated.
Select menu item
to activate it.
13.5 Zero Range
The “Zero Range” value is used in some of the application functions as a reference to identify whether
the sample is loaded.
If the display is within the Zero Range, the balance determines that a sample is not loaded. If the dis-
play goes more than five times the Zero Range it determines that a sample is loaded. Application func-
tions which involves judgment of whether a sample is loaded or not work in accordance with the Zero
Range setting.
Set the “Zero Range” value in menu item
using the number of counts displayed in gram-display.
The setting range is 01 to 99 with 01 being the default value. Even when weighing will be done in
another unit, Zero Range setting is made by only gram value.
Notes
• During continuous output, the Communication symbol may appear to
remain lit. If the transfer speed of the data output is slow, the display
may flash. Increase the transfer speed as much as possible and set the
handshake off (menu item
).
• Pretaring Value (
) function cannot be used with Auto Print functions.