61
14. Application Functions
Read Chapter 7 Before Use of
These Pages
Notes
• Standby state is not available in the Animal Weighing mode.
• Press the
[POWER]
key to initiate the power supply standby state.
• On the premise of weighing animated objects, the stability detection
band is automatically extended in the Animal Weighing mode. Repro-
ducibility of the measurement data is slightly less than with other modes.
• When the animal being weighed cannot be controlled and the automatic
printing does not occur, press the
[PRINT]
key to output the displayed
value. Then unload the animal. Even if the stability symbol lights before
the animal is removed, data is not printed again.
• By setting a larger stability detection band in the menu, the stability sym-
bol will light more readily.
• If the balance is slow to return to the zero point, set a larger Zero Range
value.
• Pretaring Value (13.7) function cannot be used with Animal Weighng
function.