V
Follow operation instruction
Correctly set draft shield, sample pan supporter, sample pan handler and sample pan.
Hold the handle of heater lid when opening or closing heater lid.
Use sample pan handler when removing sample pan.
Never touch any metal parts of heater unit and surrounding parts when removing sample pan.
Cool the unit down to ambient temperature in a safe location after measurement.
Always make sure that the unit has cooled down sufficiently before covering it with a dust cover.
Warning on high temperature
The shadowed parts become especially hot during measurement. Only handle the parts marked with circles when
operating.
Do not measure hazardous samples
Use the Electronic Moisture Balance only for measuring the moisture content in a sample by moisture evaporation
through the heating process.
Set the drying temperature within the safe temperature range of each sample.
The heater inside the Electronic Moisture Balance reaches temperatures higher than the preset sample-heating
temperature.
The surface of the quartz tube of the heater reaches approximately 500°C (900°F) even during normal
measurements. In addition, the nickel-chromium resistance wire inside the heater reaches approximately 900°C
(1700°F) and is not airtight. Measurements of combustible or flammable samples pose a risk of fire.
SALES & TECHNICAL SUPPORT
QAQC LAB www.qclabequipment.com TEL (804) 318-3686
Содержание MOC-120H
Страница 2: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 10: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 75: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 76: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...