39
10. Error Messages
If any of the following error messages is displayed, follow the procedures described below to
check for the cause and take appropriate action.
If the error cannot be removed, contact your Shimadzu representative.
Message
Er301
Er302
Er303
Er305
Er306
Er601
Er602
Er603
Er802
Description
Temperature sensor failure
Heater overheated
This error indicates the existence of an extremely hazardous condition. Turn the power off
immediately and contact your Shimadzu representative.
Heater control failure
Heater lid is open during measurement (for 15 sec.).
Lift mechanism failure
Internal clock failure
Action to be taken
Remove protective sheet located under the pan supporter
column. (See item
3
in “6. Assembly and Installation of Main
Unit” on p. 9 for instructions.)
The minimum weight which may be measured by the MOC-120H
is 0.5 grams. Measure again using a sample with a weight of 0.5
grams or greater. Press the
TARE/RESET
key to make the error
message disappear.
The maximum weight which may be measured by the MOC-120H
is 120 grams. Measure again using a sample with a weight of
120 grams or smaller. Press the
TARE/RESET
key to make the error
message disappear.
This error is displayed when a sample is added during
measurement. Press the
TARE/RESET
key to make the error message
disappear.
Press the
TARE/RESET
key to make the error message disappear.
Turn off the power and then turn the power on again.
Use a calibration weight of the correct weight value.
Perform balance calibration again with the MOC-120H placed on
a flat, stable surface not subject to the effects of external
vibration, drafts, or breezes.
Turn the power off, make sure that the power conversion switch
located on the back is set to the correct position.
Message
Er102
Er103
Er104
Er201
Er202
Er401
Er501
Er502
Er701
Description
The protective sheet has not
been removed.
The sample is too light (i.e.,
under 0.5 grams).
The sample is too heavy (i.e.,
over 120 grams).
Invalid moisture content value
(Occurs when weight increases
by more than 0.1 grams)
Invalid moisture content value
(Occurs when weight being
measured is less than –1 grams)
Balance communications error
Invalid calibration weight
Instability during balance
calibration
Power error
If any of the following errors are displayed, contact your Shimadzu representative.
SALES & TECHNICAL SUPPORT
QAQC LAB www.qclabequipment.com TEL (804) 318-3686
Содержание MOC-120H
Страница 2: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 10: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 75: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 76: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...