![Shimadzu MOC-120H Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/shimadzu/moc-120h/moc-120h_instruction-manual_1256928055.webp)
43
12-2 Procedure for Obtaining Predicted Measurements
1) Use preparatory measuring mode to obtain the appropriate bias value for the predictive mea-
surement. (See “12-2-1 Calculating the bias to be specified when operating in predictive measure-
ment mode” for instructions.)
2) Use the bias obtained in preparatory measuring mode in step 1) to perform the measurement,
and evaluate the predicted measurement. (See “12-2-2 Evaluating predicted measurements”
for instructions.)
3) Measure in predictive measuring mode. (See “12-2-3 Preforming predictive measurements” for
instructions.)
12-2-1 Calculating the bias to be specified when operating in predictive measurement mode
Display
Operating instructions
Set the mode to preparatory measuring mode. (See 8-2-2 “6) Specifying
settings for preparatory measuring mode” for instructions.)
Specify the desired predicted value convergence range. In the example shown
here, a range of 0.5% is specified.
•
Note that the predicted value convergence range specified affects the amount of
time required for measurement and the accuracy of predictive measurements,
and that you should specify a range according to your objectives.
Specify the automatic halting conditions to be used (e.g., 0.05%).
Specify the drying temperature to be used. (For further information, see “8-2-3
TEMP (Drying temperature)”.) In the example shown here, a temperature of
120°C is specified.
Set the bias to 0.00%. (See “8-2-4 Bias” for further information.)
Place the sample onto the sample pan and begin measurement. (In the
example shown here, the initial weight is 10.478 grams.)
When the final predicted value has been calculated, it will be displayed. (In the
example shown here, the measuring time was 7 minutes and the predicted
measurement is 15.8 percent.)
Step
1
2
3
4
5
6
7
. . . . .
. . . . .
SALES & TECHNICAL SUPPORT
QAQC LAB www.qclabequipment.com TEL (804) 318-3686
Содержание MOC-120H
Страница 2: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 10: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 75: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 76: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...