![Shimadzu MOC-120H Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/shimadzu/moc-120h/moc-120h_instruction-manual_1256928039.webp)
27
Specifying settings for the TEMP (Drying temperature)
8-3-3 TEMP (Drying temperature) Setting
The following describes the procedure to specify the drying temperature in conducting
measurements.
While the default drying temperature is 120°C, depending on the type of sample, its moisture content,
or other conditions the proper drying temperature may vary. To find the proper drying temperatures
for different types of materials, conduct repeated measurements until you find the correct
temperature to use for each material.
The drying temperature may be set within the range from 30°C to 200°C (from 30°C to 180°C in 'TT'
mode) in 1-degree increments.
* Note that the drying temperature refers not to the temperature of the sample but the temperature of the
temperature sensor. The actual temperature of samples during drying may vary depending on the color
of the sample, its moisture content, or the type or form of the sample.
* Generally speaking, drying is performed more quickly when a higher drying temperature is specified,
but if the temperature specified is too high it might result in the sample becoming burned, thus making
it impossible to obtain accurate measurements.
Display
Key(s) used
Operating instructions
With the display of the weight in grams, press the
CONDITION
key.
If a password has been specified, the message ‘PASS’ will be
displayed to indicate that the password should be entered. Use
the
key,
key and
ENTER
key to enter the password. Follow
the instructions under step
7
in “9-2-7 Specifying a password” to
set or delete a password.
The
‘CONDITIONS’
display area will begin to flash. Press the
SELECT
key until
‘TEMP’
begins to flash
With the
‘TEMP’
indicator flashing, press the
ENTER
key. This will
cause control to shift to specification of the drying temperature
and the currently specified drying temperature to begin to flash.
Press the
key or
key to select the desired drying
temperature. Note that here a temperature of 105°C has been
specified.
When the desired drying temperature flashes, press the
ENTER
key.
The display of the specified temperature will change from a
flashing to a constantly lit display and the
‘TEMP’
indicator will
begin to flash instead.
If you wish to specify other measuring
conditions
settings, press
the
SELECT
key to select the menu.
To exit from specifying measuring
conditions
settings, press the
CONDITION
key instead.
Step
1
2
3
4
5
6
7
CONDITION
ENTER
SELECT
ENTER
SELECT
CONDITION
SALES & TECHNICAL SUPPORT
QAQC LAB www.qclabequipment.com TEL (804) 318-3686
Содержание MOC-120H
Страница 2: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 10: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 75: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...
Страница 76: ...SALES TECHNICAL SUPPORT QAQC LAB www qclabequipment com TEL 804 318 3686...